Calibration Thermometer
QUV Thermometer Calibration
The QUV black panel temperature sensor needs to be calibrated periodically by the user to assure accurate and consistent results. Calibrating the QUV black panel temperature sensor is simply performed with any standard reference thermometer.Temperature sensors do not require normal replacement. In the event of an unlikely failure, a replacement can be ordered under the following part number:
- V-2156-X QUV Black Panel Temperature Sensor (replace as necessary)
Q-SUN Thermometer Calibration
All Q-SUN black panel temperature sensors need to be calibrated periodically by the user to assure accurate and consistent results. Calibrating the black panel temperature sensor is simple using the Universal Calibrator system’s UC202 calibrated temperature sensor and UC1 handheld display, or the historical CT202 thermometer, and takes only a few minutes. Note that CT202 and UC202 thermometers comes with either an uninsulated black panel or insulated black panel sensor, and must match the type of sensor actually used in the Q-SUN tester.
Calibration of the UC202 and CT202 thermometers needs to be performed annually. The UC202 thermometers were designed to be disposable and cost-efficient, and should be replaced annually with a freshly-calibrated UC202. The CT202 thermometer needs to be returned to Q-Lab once per year for an inexpensive recalibration. UC202 units may also be returned for calibration for users who prefer recalibration to replacement. Our calibration labs are ISO 17025 accredited by the American Association of Laboratory Accreditation (A2LA - Ohio Certificate Number 2382-01, China Certificate Number 2383-02 and England Certificate Number 2383-03). Calibrations are traceable to the U.S. National Institute of Standards and Technology (NIST). See Recalibration Return Procedure.
In the Q-SUN Xe-1, the tester can have an optional chamber air temperature sensor (included for a model Xe-1 with chiller). Calibrating the Xe-1 chamber air temperature sensor is simply performed with any standard reference thermometer.In both the Q-SUN Xe-2 and Xe-3 models, the low-cost, combination relative humidity/chamber air temperature sensor was thoughtfully designed to be disposable. It should be replaced once a year and cannot be recalibrated.
Common Q-SUN thermometer part numbers are shown below:
- X-10235-X Q-SUN Xe-1/Xe-3 Black Panel Temperature Sensor (replace as necessary)
- X-7778-X Q-SUN Xe-1/Xe-3 Insulated Black Panel Temperature Sensor (replace as necessary)
- XR-11463-X Q-SUN Xe-2 Black Panel Temperature Sensor (replace as necessary)
- XR-11464-X Q-SUN Xe-2 Insulated Black Panel Temperature Sensor (replace as necessary)
- X-10415-K Q-SUN Relative Humidity/Chamber Air Sensor (replace annually)
- UC1 Universal Calibrator Handheld Display
- UC202/BP Universal Calibrator System Q-SUN Black Panel Calibration Thermometer
- UC202/IBP Universal Calibrator System Q-SUN Insulated Black Panel Calibration Thermometer
- UC202/RECAL Recalibration of UC202 Calibration Thermometer (annual replacement is recommended but annual recalibration is available)
- CT202/RECAL Recalibration of CT202 Thermometer (recommended annually)